Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G4267-40033 | Transport Protection |
Prerequisites
|
NOTE
The graphics are exemplary. The procedure shown is the same for your module.
Open the front door.
Manually move the robot arm to the right back corner of the lobby (park position).
Carefully slide the right protection foam piece into the sampler.
Position the foam next to the hydraulic box such that it sits behind the needle port.
Slide the left protection foam piece into the sampler.
Firmly push it into place until it snaps in behind the hotel support frame.
Place the protection foam of the hydraulic box in a way that it snaps onto the analytical head.
Verify that all three protection foams are in the correct position.
Close the front door.