| For biocompatible modules use bio / biocompatible parts only! Do not mix with bio-inert parts. |
When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G4767-60500 | Calibrated Sample Loop, 100 µL, right (blue coded) (for G7167C) OR |
1 | ![]() | G7137-60500 | Sample Loop MP35N 100 µL, right (blue/orange coded) (for G7137B) |
Prerequisites
|
WARNING
Risk of injury by uncovered needle
An uncovered needle is a risk of harm to the operator.
CAUTION
Mismatching sample loop configuration
Damage to the system
CAUTION
Blockages inside of the needle assembly union
Use a 1/4-inch wrench to tighten the fitting of the loop capillary.
Close the front door.
In the Local Controller close Change loop.
OR
In the Agilent Lab Advisor software Change Loop, click Back and wait until the needle is in the needle park station.
NOTE
Choose the Change Needle procedure, if you need an auto referencing step included.
NOTE
If you have changed the sample loop, verify that the correct sample loop is configured in the CDS, see Setting Up the Autosampler With OpenLab ChemStation.
For G7167C and G7137B, proceed with the subsequent steps.
Perform Hydraulic Path Leak Test.
Perform Sampler Leak Test if needed for troubleshooting in case of Hydraulic Path Leak failure.