When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]() | 5182-3466 | Torx screwdriver T10 |
1 | ![]() | 5023-3138 | Reversible Screwdriver + Blade 1,0 x 5,5 |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G7129-87017 | Seat Assembly, PEEK, 0.17 mm, 1260 Vialsampler OR |
1 | ![]() | G7129-87012 | Seat Assembly, PEEK, 0.12 mm, 1290 Vialsampler OR |
1 | ![]() | G7129-87117 | Seat Assembly, Vespel, 0.17 mm, 1260 Vialsampler OR |
1 | ![]() | G7129-87112 | Seat Assembly, Vespel, 0.12 mm, 1290 Vialsampler |
Prerequisites
|
WARNING
Risk of injury by uncovered needle
An uncovered needle is a risk of harm to the operator.
NOTE
When the instrument setup has changed, configure the new setup in the online view of the Chromatographic Data System. See The Vialsampler User Interface.
NOTE
When the instrument setup has changed, configure the new setup in the online view of the Chromatographic Data System. See The Vialsampler user interface.
NOTE
It is recommended to always exchange the needle assembly and the needle seat at the same time to prevent premature leakage.
NOTE
The use of a Vespel needle seat (G7129-87112 and G7129-87117) limits the maximum system pressure to 600 bar.
In the Agilent Lab Advisor software select Service & Diagnostics > Maintenance Positions > Change Needle, Loop and Seat , click Start and wait until the needle assembly is in maintenance position.
OR: In the Local Controller start the maintenance mode and select Change Needle, Loop and Seat function.
In Lab Advisor use Next to move the needle in the down position.
Disconnect the seat capillary fitting from the injection valve (port 5) and ease out the needle seat from the leak tray with the help of a flathead screwdriver.
Insert the new needle seat and press it firmly into position. Then, connect the seat capillary fitting to port 5 of the injection valve.
Align the needle tip with the seat and ensure that they are concentric. If needed, carefully bend the needle into the right position with your fingers.
In Lab Advisor use Next to lift the needle slowly into the up position.
In Lab Advisor, select Next to move the needle into the needle seat and then Back to finish the maintenance procedure.
OR: In the Local Controller, select Continue to move the needle into the needle seat and then Exit to finish the maintenance procedure.
Perform a System Pressure Test.