When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]() | 5023-2524 | Hex Key Set (part of the G7120-62708 InfinityLab LC Series Tool Kit) |
1 | ![]() | 01018-23702 | Insert tool (for 100 µL seals) |
1 | ![]() | G4226-43800 | Seal insert tool, OR (for 40 µL seals) |
1 | ![]()
|
| Cleaning tissue and appropriate solvent like isopropanol or methanol |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 0905-1719 | PE Seal |
1 | ![]() | 5067-5678 | Piston, 100 µL, Zirconium oxide Replacement only necessary in case of wear (Optional) |
Prerequisites
|
Install the new metering seal using the plastic side of the insert tool. Press it firmly into position. Avoid any offset angle as it might deform the seal.
In the Local Controller exit the maintenance mode and select Change metering device function.
OR
In Agilent Lab Advisor software system screen exit Service & Diagnostics (Tools) > Maintenance Positions > Change Metering Device click End and wait until the metering device is in Home position.
Perform a Pressure Test.