When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]() | 8710-2392 | 4 mm Hex key |
1 | ![]() | 01018-23702 | Insert tool, OR |
1 | ![]() | G4226-43800 | Seal insert tool |
In the Local Controller start the maintenance mode and select Service Position.
OR
In the Agilent Lab Advisor software select Service & Diagnostics in the system screen (Tools) > Maintenance Positions > Change Metering Device , click start and wait until the metering device is in maintenance position.
Take the metering device. Push against the rear side of the metering device and rotate a quarter left to release the bayonet lock.
If dirty: Clean the piston with an appropriate solvent. If scratched: Replace the piston by a new one.
Carefully remove the metering seal using the steel side of the insert tool. Clean the chamber with an appropriate solvent and ensure that all particulate matter is removed.