When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]() | 5023-2524 | Hex Key Set (part of the G7120-62708 InfinityLab LC Series Tool Kit) |
1 | ![]() | 01018-23702 | Insert tool (for 100 µL analytical head) |
1 | ![]() | G4226-43800 | Seal insert tool (for 40 µL analytical head) |
1 | ![]()
|
| Pair of tweezers |
1 | ![]() | 05980-60051 | Cloths, clean, lint-free 15/pk |
1 | ![]()
|
| Isopropanol or any other appropriate solvent |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 0905-1717 | Metering Seal, 40 µL OR |
1 | ![]() | 0905-1503 | Metering Seal, 100 µL OR |
1 | ![]() | 0905-1294 | Metering Seal, 900 µL |
1 | ![]() | 5067-5920 | Piston, 40 µL, Zirconium oxide for 40 µL analytical head OR |
1 | ![]() | 5067-5678 | Piston, 100 µL, Zirconium oxide for 100 µL analytical head OR |
1 | ![]() | G4267-60462 | Piston, 900 µL, Sapphire |
Prerequisites
|
NOTE
When the instrument setup has changed, configure the new setup in the online view of the Chromatographic Data System. See The Vialsampler user interface.
In the Agilent Lab Advisor software, select Service & Diagnostics > Maintenance Positions > Change Metering Seal and Piston , click Start and wait until the metering plunger is in maintenance position.
OR: In the Local Controller start the maintenance mode and select Change Metering Device function.
Use a seal insert tool to carefully remove the seal.
Clean the chamber and the plunger with a piece of lint-free cloth and an appropriate solvent, such as isopropanol. Check the plunger for sings of scratches or damage and if necessary, replace it. Ensure that all particulate matter is removed from the chamber.
Use the plastic side of the insert tool to install the new seal. Press the seal firmly into position.
In Lab Advisor, select Next to move the piston back into its normal position and then Back to finish the maintenance procedure.
OR: In the Local Controller, select Continue to move the piston back into its normal position and then Exit to finish the maintenance procedure.
Perform a System Pressure Test.