Error ID:
Intensity Test Evaluation
NOTE
If only one range fails and the application does not require this range, the lamp may not be changed.
Probable cause | Suggested actions | |
---|---|---|
1 | Absorbing solvent or air bubble in flow cell. | Ensure the flow cell is filled with water, and free from air bubbles. Repeat test without flow cell and compare results. |
2 | Incorrect calibration | Recalibrate and repeat the test. |
3 | Dirty or contaminated flow cell. |
Run the cell test. If the test fails, flush the flow cell. See also Clean the Max-Light Cartridge Cell. Run the cell test. If the test fails, flush the flow cell. See also Maintenance of Standard, Semi-Micro or Micro Flow Cell. |
4 | Dirty or contaminated optical components. | Please contact your Agilent service representative. |
5 | Old UV-lamp. | Exchange the UV lamp. |
6 | Defect optical unit. | If the test fails with flow cell and new UV-lamp, please contact your Agilent service representative. If the test fails with new UV-lamp, please contact your Agilent service representative. |