The test determines the leak rate of the system between pump outlet valves and a blank nut. The blank nut can be positioned at different locations in the system before the flow cell, to determine and verify the leak rate of individual modules and components. The test allows for setting the pressure at which the test is performed. The leak rate of high pressure parts is not always a linear function and therefore it is recommended to perform the test at a pressure that corresponds to the normal operating pressure of the system.
适用于
|
所需部件
数量 | 部件号 | 描述 | |
---|---|---|---|
1 | ![]() | 5067-6127 | 堵头 SL For 1260 and 1290 Infinity II systems. OR |
1 | ![]() | 5043-0277 | 生物相容设备的 PEEK 堵头 For Bio-inert or 1290 Bio systems. |
Run the System pressure test with the Agilent Lab Advisor (for further information, see Online-Help of user interface).
base-id: 6301375755
id: 9007205556116747