适用于
|
所需工具
数量 | 部件号 | 描述 | |
---|---|---|---|
1 | ![]()
|
| 实验室天平 |
所需部件
数量 | 部件号 | 描述 | |
---|---|---|---|
1 | ![]()
|
| DAB/Ph Eur/BP/JP/NF/USP 级蔗糖 |
1 | ![]() | 5190-1539 | Syringe 5mL with Luer-Lock |
1 | ![]() | 9301-0407 | 进样针、外部阀接头、SST |
1 | ![]() | 5190-5111 | 进样针过滤器,0.45 µm,100 个/包 |
1 | ![]() | 0100-1516 | 手紧接头,PEEK,2/pk |
Preparation of the sucrose calibration solution (final concentration: 3.5 g/L).
Wait five minutes and shake again. The solution is now ready for use.
Preparing the pump.
Flush the degasser and pump.
Run the RID Tools Screen with the Agilent Lab Advisor (for further information see Online-Help of user interface). (If an Agilent pump is part of the system, the pump section is active.)
Purging the sample and reference cells.
Fill the sample cell with calibration solution.
注意
Do not inject the calibration solution without the sample filter.
Leave the RID Tools Screen.