This procedure shows the installation of the Agilent 1260 Infinity III Manual Injector (G1328C) as an example. The procedure is the same for the Agilent 1260 Infinity III Bio-Inert Manual Injector (G5628A)
小心
“到货即损”问题
如果有损坏迹象,请不要尝试安装模块。需要由 Agilent 进行检查,以评估仪器是否状况良好或已损坏。
注意
The manual injector can be installed at the left- or right-hand side of the instrument stack.
Place the baseplate on the bench.
Place the Base Plate Adaptor on the connectors of the organizer plate.
Connect the organizer plates with the base plate.
Place (1.) and screw (2.) the mounting pole into one of the two holes in the organizer plate.
Slide the manual injector onto the mounting pole (1.) and tighten the mounting screw (2.).
Install other system modules on top of the manual injector baseplate (see following figure).
Connect the capillaries to the manual injector (see Flow Connections (G1328C) ).