Running the Pump Elasticity Calibration from the Agilent Lab Advisor Software
适用于
|
所需工具
数量 | 部件号 | 描述 | |
---|---|---|---|
1 | ![]() | 8710-0510 | 开口扳手 1/4-5/16 英寸 |
所需部件
数量 | 部件号 | 描述 | |
---|---|---|---|
1 | ![]() | G1312-67500 | 限流毛细管 |
前提条件
|
注意
Make absolutely sure that the pump is very thoroughly flushed with the solvent to be calibrated before starting the procedure! Any trace of other solvents or the smallest air bubble inside the flow path definitely will cause the calibration to fail!
注意
If a solvent selection valve is installed flush all four solvent channels to avoid that air from a dry solvent intake tube is drawn into the flow path upon initialization.
从测试选择菜单中选择泵弹性校准。
开始测试并按照说明进行操作。
注意
请确保完成测试后打开冲洗阀以释放压力。否则泵可能产生超压故障。