The ASTM noise test determines the detector noise over a period of 20 minutes. The test is done with installed Max-Light Cartridge Cell or Max-Light Cartridge Test Cell.
This test does also check for the drift. It is also part of the “Self Test” (without checking for the drift).
If the test is performed with the Max-Light Cartridge Test Cell, the test results are not influenced by solvent or pump effects.
适用于
|
所需部件
数量 | 部件号 | 描述 | |
---|---|---|---|
1 | ![]()
|
| 最大光强卡套流通池(充满水) |
1 | ![]()
|
| 最大光强卡套测试池 |
前提条件
|
Run the ASTM Drift and Noise Test with Agilent Lab Advisor (for further information see Online-Help of user interface).
base-id: 3526336907