Test evaluation
注意
If the test fails with Flow Cell and new UV-lamp, the optical unit must be checked/replaced.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Absorbing solvent or air bubble in Flow Cell. | Repeat calibration with Flow Cell and compare results. |
2 | Dirty or contaminated Flow Cell. | 确保流通池充满水,没有气泡。 Recalibrate. |
3 | 紫外灯很旧。 | 更换紫外灯。 |
4 | Dirty or contaminated optical components. | Run the Cell Test. If the test fails, flush the flow cell. |
base-id: 3358836619