Holmium Oxide Test evaluation
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Absorbing solvent or air bubble in flow cell. | 确保流通池充满水,没有气泡。 |
2 | Incorrect calibration | Recalibrate (see Wavelength Calibration ) and repeat the test. |
3 | Dirty or contaminated flow cell. | Run the cell test (see Cell Test ). If the test fails, exchange the flow cell windows. |
4 | Dirty or contaminated optical components (achromat, windows). | Clean optical components with alcohol and lint-free cloth or replace the parts (see Intensity Test ). |
5 | Old or non-Agilent lamp. | 更换紫外灯。 |