小心
“到货即损”问题
如果有损坏迹象,请不要尝试安装模块。需要由 Agilent 进行检查,以评估仪器是否状况良好或已损坏。
注意
The Installation of the leak tray and the external leak sensor is optional.
If you want to attach the leak tray to the valve drive you need to slide the leak tray on to the valve drive before having both clamps attached to the valve drive.
Ensure the power switch at the rear of the module is OFF and the power connector is unplugged.
Identify the marks for the screws on the side panel of the cabinet assy and screw the valve rail to the side panel of the cabinet.
Slide the first clamp on to one side of the valve drive housing.
Install the leak tray and leak sensor to the valve.
Slide the second clamp on to the same side as the first one.
Slide the valve on to the valve rail.
Attach the waste tubing to the leak plane and guide it to a proper waste container.
Connect the power cable to the power connector at the rear of the module.
Connect the CAN interface connection.
Power on the module by switching the Power button at the rear of the module if a valve head has been already installed. Otherwise continue with G7116B_Installation of the Valve Heads .
base-id: 3181722507
id: 3181722507