The Agilent 1260 Infinity III SFC Control Module is heavy (approximately 26 kg or 56 lbs). Enlist the aid of a co-worker to share the lifting load in order to avoid possible injury. Select the laboratory bench space before your system arrives. Pay special attention to the total height requirements. Avoid bench space with overhanging shelves. Pay special attention to the total weight of the modules and solvents you have in addition to the G4301A SFC Control Module. Make sure that your laboratory bench can support this weight.
类型 | 规格 | 注释 |
---|---|---|
重量 | 26 kg(56 磅) |
|
尺寸(高 × 宽 × 长) | 60 cm x 26 cm x 48 cm |
|
线路电压 | 100 - 240 VAC, ±10% |
|
线路频率 | 50 - 60 Hz, ±5% |
|
耗电量 | 700 VA 最大值 |
|
操作温度 | 15 至 30°C(59 至 86°F) |
|
非操作温度 | -40 至 70°C(-40 至 158°F) |
|
湿度 | 温度为 25–40°C 时,<95% | 无冷凝 |
实验室通风 | 实验室通风需要最少 6 次空气对流/小时 | CO₂ 检测,建议达到 5000 ppm 时发出警报 |
排气口容量 | >20 L/min 时具有持续负压 |
|
操作海拔高度 | 最高 3000 m (9842 ft) |
|
安全标准:IEC、EN、CSA、UL | 过电压类别 II,污染程度 2 | 仅限室内使用 |
ISM 分类 | ISM 组 1 类别 A | 根据 CISPR 11 |
类型 | 规格 |
---|---|
入口 CO₂ 主体纯度 | > 99.99% 的蒸汽;> 99.999% 的液体 |
入口 CO₂ 相态 | 非浸管高压气瓶中为气态;商用 CO₂ 输送系统中为液态 |
入口 CO₂ 供应压力 | 40 - 70 bar (580–1000 psi) |
入口 CO₂ 温度 | 15 至 30°C (59 - 86°F) |
液态制冷剂 | 含 30% 丙二醇的去离子水;专有抗氧化剂;为确保安全,添加了红色染料 |
制冷剂体积 | < 280 mL |
类型 | 规格 |
---|---|
高压流路 | 300 和 400 系列不锈钢、碳填充 PEEK、UHMW PE、蓝宝石、陶瓷、金 |
低压流路 [废液、冲洗泵、漏液托盘] Only when a legacy wash pump is installed | 316 不锈钢、PEEK PTFE、FEP、CTFE CPE;LDPE Tygon PVC |
蒸汽废气 | Tygon PVC |
[1]仅当安装了传统冲洗泵时。
类型 | 规格 |
---|---|
液压系统 | 带专用马达控制的单柱塞杆 |
总液压体积 | < 5 mL(在压力小于 70 bar 时) < 25 mL(在压力达到 400 bar 时) |
冷却器系统 | 带二次空气/液体冷却回路的热电冷却 |
冷却器温度 | -20 至 9°C(-4 至 48.2°F) |
背压调节 (BPR) 系统 | 带专用驱动控制的低体积膜片类型;可更换的 BPR 头组件;更换头后无需重新校正 |
增压泵压力范围 | 流量为 5 mL/min 时,100–400 bar(按需) |
BPR 热范围 | 40 至 70°C(104 至 158°F) |
BPR 热精度 | ±1 K |
BPR 可设置压力范围 | 100 – 400 bar |
背压精度 | ±0.25 bar |
背压热精度 | ±1 K |
可编程背压梯度 | 是。最多 20 个时间表条目。 |
通讯 | USB 2.0;APG 远程:就绪、启动、停止和关机信号;继电器触点关闭(仅清洗泵) |