If the Assist Hub is connected to the Internet and has an Assist Control Software Version of 2.0 or higher, then the software generates a notification when an update is available.
You can manually check for software updates. Updates are available under:https://update.pl29.agilent.com/infinitylab. Check regularly for new updates.
You can use a USB storage media to store the update files. It must be of type exFAT, FAT32, EXT4. Both, the Assist Interface and the Assist Hub consist of a USB connector.
If you start InfinityLab Assist remotely in a Web browser, you can also store the update files to a known location on the PC and select software or firmware updates manually. A Chromium-based browser is required to open the Assist Control Software.
从主工具栏中选择 。
将显示“设置”屏幕。“常规”选项上的红点表示更新可用。
在“常规”选项卡上,选择更新。
“更新”向导将启动。如果自动检查更新处于活动状态并且可以访问源,则将列出可用的软件和固件更新。
否则,选择检查更新。
如果可以访问更新源,则将列出可用的软件和固件更新。
如果您在网页浏览器中远程启动 InfinityLab Assist,也可以选择浏览,自己搜索文件系统并手动选择软件或固件更新。
选择所需的更新。
选择使用所选继续。
如果更新向导无法打开文件,则需要再次检查是否存在更新。否则,将显示更新文件的内容。
如果新固件版本与已安装的 LC/CE 驱动程序版本不兼容,更新向导将显示所需 LC/CE 驱动程序版本的版本号。要跳过固件更新,请清除固件更新复选框。或安装所需的 LC/CE 驱动程序版本。或取消更新向导。
要筛选出不应用于您仪器中模块的更改,请选中仅硬件配置相关的更改复选框。
要显示发布说明,选择 。
选中更新后自动重新启动复选框。
选择安装更新。
更新向导将上传更新文件。进度指示器显示进度。上传成功完成后,Assist Hub 将在 60 秒后重新启动。如果仪器当时正忙,则将在仪器再次空闲时执行重新启动。
将显示稍后重新启动和现在重启按钮。现在重启按钮只有在仪器空闲时才处于活动状态。
要推迟重启,请选择稍后重新启动。
在 Assist Hub 关闭并再次打开后,Assist Hub 将使用已安装的更新重新启动。
要立即重启,请选择现在重启。
Assist Hub 将使用已安装的更新重新启动。
清除更新后自动重新启动复选框。
选择安装更新。
更新向导将上传更新文件。进度指示器显示进度。当上传成功完成且仪器处于空闲状态时,将显示现在重启按钮。如果仪器正忙,将推迟重新启动,直到仪器处于空闲状态。
要立即重启,请选择现在重启。
Assist Hub 将使用已安装的更新重新启动。
base-id: 13027880459