Ferramentas necessárias
Quant. | p/n | Descrição | |
|---|---|---|---|
1 | ![]() | 0100-1710 | Mounting Tool for Tubing Connections |
1 | ![]() | 8710-0899 | Screwdriver Pozidrive Shaft |
Peças necessárias
Quant. | p/n | Descrição | |
|---|---|---|---|
1 | ![]() | G1311-67701 | Multi channel gradient valve (MCGV) |
Pré-requisitos
|
NOTA
For best performance and life time, use lower channels A and D for aqueous solvents in buffer applications.
Use the mounting tool for removing tubing connections between the degassing unit and the MCGV.
Remove the inlet weaver, see Replace the Inlet Weaver.
Remove the cover from the MCGV.
Disconnect the MCGV cable (1), unscrew the two screws (2) and remove the valve (3).
Place the new MCGV into position.
NOTA
Make sure that channel A of the MCGV is put at the bottom-right position.
Tighten the two screws (1) and connect the cable to its connector (2).
Install the MCGV cover.
Install the inlet weaver, see Replace the Inlet Weaver.
Reconnect solvent tubes for channels A-D from the MCGV to the degasser outlets.