Para módulos bio-inert, utilize somente peças bio-inert! |
| Para módulos biocompatíveis, utilize somente peças bio/biocompatíveis! Não misture com peças bio-inert. |
Quando
|
Ferramentas necessárias
Quant. | p/n | Descrição | |
|---|---|---|---|
1 | ![]() | 8710-0510 | Chave de boca 1/4-5/16 de polegada |
Peças necessárias
Quant. | p/n | Descrição | |
|---|---|---|---|
1 | ![]() | G4267-60300 | Sample Loop Flex 20 µL, right (red coded) |
1 | ![]() | G4267-60400 | Sample Loop Flex 40 µL, right (green coded) |
1 | ![]() | G4267-60500 | Sample Loop Flex 100 µL, right (blue coded) |
1 | ![]() | G7167-68500 | Sample Loop Cartridge 500 µL right |
1 | ![]() | G7167-68900 | Sample Loop Cartridge 900 µL right |
1 | ![]() | G5668-60500 | Bio-inert Sample Loop 100 µL (for G5668A) |
1 | ![]() | G7137-60300 | Sample Loop MP35N 20 µL, right (red/orange coded) |
1 | ![]() | G7137-60400 | Sample Loop MP35N 40 µL, right (green/orange coded) |
1 | ![]() | G7137-60500 | Loop de amostra MP35N 100 µL, direito (código azul/laranja) |
Further sample loops for the Dual-Needle option are available, see Sample Loops and Capillaries (Dual Needle).
Pré-requisitos
|
ADVERTÊNCIA
Risco de ferimento por agulha desencapada.
Uma agulha desencapada pode ferir o operador.
CUIDADO
Divergência de configuração de loop de amostra
Danos ao sistema
NOTA
If you have changed the sample loop, verify that the correct sample loop is configured in the CDS (see Setting Up the Autosampler With OpenLab ChemStation).
NOTA
For details on the setup of the dual-needle system, see Modify Capillaries.
CUIDADO
Blockages inside of the needle assembly union
Use a 1/4-inch wrench to tighten the fitting of the loop capillary.
Install the shorter capillary of the sample loop cartridge to the analytical head.
Pinch and reinsert the needle assembly and the connected sample loop capillary into the z- arm coupler.
NOTA
Check the tension of the loop capillary. This must be forced and guided to the hydraulic box to prevent it from being caught by the Z-drive.
In the Local Controller close Change needle /seat.
OR
In Agilent Lab Advisor software Change needle/loop. Click NEXT and wait until the needle is in the needle park station.
Click Back to leave the Maintenance window.
NOTA
If you need an autoreferencing step included, you must choose the change needle procedure.
NOTA
If you have changed the sample loop, verify that the correct sample loop is configured in the CDS (see Setting Up the Autosampler With OpenLab ChemStation).