É necessário realizar inspeção periódica do instrumento para garantir que ele seja usado com segurança. Representantes de serviço da Agilent podem realizar essas inspeções periódicas mediante contrato. Para obter informações sobre o contrato de inspeção de manutenção, entre em contato com seu representante da Agilent.
The following pages describe maintenance (simple repairs) of the detector that can be carried out without opening the main cover.
Procedure | Typical Frequency | Notes |
|---|---|---|
Flow cell exchange | If application requires a different flow cell type or if defective. | Complete Assembly A wavelength calibration check should be performed after replacement.
If the flow cell is removed and inserted, then a quick calibration check is performed. If this fails, you must do a wavelength recalibration, see Wavelength Verification and Calibration. |
Flow cell flushing | If flow cell is contaminated. |
|
Leak sensor drying | If leak has occurred. | Check for leaks. |
Leak handling System replacement | If broken or corroded. | Check for leaks. |