Quando
|
Ferramentas necessárias
Quant. | p/n | Descrição | |
|---|---|---|---|
1 | ![]() | 5023-3138 | Chave de fenda reversível + lâmina 1,0 x 5,5 |
Peças necessárias
Quant. | p/n | Descrição | |
|---|---|---|---|
2 | ![]() | 2110-1486 | Fuse 2 AT250 V |
ADVERTÊNCIA
Electrical shock
Switch off the instrument. The power switch is located at the rear of the valve drive.
Disconnect the power cable from the power input socket.
To access the fuse drawer, gently lift the outer plastic housing of the power inlet socket using a flat screwdriver.
Pull out the fuse drawer as shown.
Replace the defect fuse(s).
Slide in the fuse drawer and push till it fits tightly.
Finally, close the fuse drawer housing, reconnect the instrument to the power line and switch it on.
base-id: 3182142219
id: 9007202436883211