Quando
|
Ferramentas necessárias
Quant. | p/n | Descrição | |
|---|---|---|---|
1 | ![]() | 8710-0510 | Chave de boca 1/4-5/16 de polegada |
1 | ![]() | 8710-1924 | Chave de boca de 14 mm |
1 | ![]()
|
| Par de pinças, OR |
1 | ![]()
|
| Palito |
Peças necessárias
Quant. | p/n | Descrição | |
|---|---|---|---|
1 | ![]()
|
| Frits |
1 | ![]()
|
| Purge Valve |
1 | ![]() | 5067-4728 | Conjunto de tampa de selo |
Pré-requisitos
|
Using a 1/4 inch wrench disconnect the pump outlet capillary from the purge valve. Disconnect the waste tube. Beware of leaking solvents due to hydrostatic pressure.
Using the 14 mm wrench, unscrew the purge valve and remove it from the purge valve holder.
Remove the seal cap from the purge valve. Using a pair of tweezers or a toothpick, remove the frit.
Place a new frit into the purge valve with the orientation of the frit as shown below (slit in frit points to the front). Reinstall the seal cap including the gold seal.
Insert the purge valve into the purge valve holder.
Tighten the purge valve and reconnect outlet capillary and waste tubing.