Quando
|
Ferramentas necessárias
Quant. | p/n | Descrição | |
|---|---|---|---|
1 | ![]() | 8710-0510 | Chave de boca 1/4-5/16 de polegada |
1 | ![]()
|
| Chave hexagonal, 4 mm |
1 | ![]() | 5023-0240 | Chave sextavada, ¼", com fendas |
Pré-requisitos
|
CUIDADO
Danos da unidade da bomba
Ligar a bomba sem a cabeça da bomba pode danificar a unidade da bomba.
NOTA
Both pump head assemblies use the same internal components. In addition, pump head A is fitted with the purge valve. The following procedure describes the removal and disassembly of pump head A (left). For pump head B (right) proceed in the same way and skip steps that deal with the purge valve.
Open the doors.
Disconnect the capillaries. Beware of leaking solvents.
Using a 4 mm hexagonal key stepwise loosen and remove the two pump head screws.
Remove the pump head assembly from the module.
base-id: 4536798475