ID do erro: 22417
Lost steps of the purge valve encoder.
Parameter: None
Causa provável | Ações sugeridas | |
|---|---|---|
1 | Purge valve drive mechanically blocked or defect. | Check installation of purge valve head. Please contact your Agilent service representative. |
base-id: 4202569483