ID do erro: 2690
Causa provável
Ações sugeridas
1
Contaminated flow cell.
Flush flow cell
2
Defective mainboard.
Please contact your Agilent service representative.
3
Defective lamp or optics.
Exchange the lamp.
base-id: 3596802955