Quando
|
Ferramentas necessárias
Quant. | p/n | Descrição | |
|---|---|---|---|
1 | ![]()
|
| Chave de fendas, Pozidriv Nº 1 PT3 |
2 | ![]()
|
| Chave de 1/4 pol. for capillary connections |
Peças necessárias
Quant. | p/n | Descrição | |
|---|---|---|---|
1 | ![]()
|
| For parts identification refer to (80 and 500). Nano Flow Cells (80 nL and 500 nL) |
Pré-requisitos
|
NOTA
For details refer to the technical note that comes with the nano-flow cell kit.
NOTA
The quartz block can be cleaned with alcohol. DO NOT touch the inlet and outlet windows at the quartz block.
Disconnect the capillaries from the capillary holder and remove the flow cell, see Remover e instalar uma Célula de fluxo.
Unscrew the cell body from the holder.
Unscrew the capillaries from the flow cell. DO NOT use the adapter at this time!
Using for example a toothpick, press on the plastic part and slide the quartz body out of the cell housing.
The quartz body and the cell seal assembly can be separated for cleaning purpose.
This figure shows the correct holding of the quartz body and the cell seal assembly.
Replace the cell seal assembly onto the quartz body. Always use a new seal assembly to exclude damage during disassembling.
Slide the quartz body completely into the cell body to the front stop (use for example a toothpick).
Insert the flow cell capillaries and tighten them fingertight. Use the wrench and torque adapter as described on Wrench plus Torque Adapter and tighten the fittings alternately.
Reassemble the flow cell body to the holder.
Install the flow cell, see Remover e instalar uma Célula de fluxo.
Perform a leak test with the flow cell outside of the detector.
If no leak is observed, install the flow cell and you are ready to work.
Make sure that the flow cell assembly is inserted correctly and fits perfectly in the optical unit (especially when PEEK capillaries are used).
base-id: 3358954251
id: 3358954251