Quando
|
ADVERTÊNCIA
Heavy weight
The module is heavy.
CUIDADO
Condensate inside
Damage to the electronics
CUIDADO
Unsecured transportation
Mechanical damage
NOTA
Moving the sampler with the Sample Thermostat installed is possible for short distances (for example, from one workbench to another). For longer transportation, remove the thermostat from the sampler and handle the units separately.
Turn off the Sample Thermostat.
Remove condensate.
NOTA
Do not tilt the module to avoid damage to the internal parts.
Remove all vials from the vial drawers.
Move the transport assembly to the park position using Instant Pilot or Lab Advisor, see Park Arm.
Turn off the sampler.
Install the Transport Protection Foam, see Install the Transport Protection Foam.
Remove the Sample Thermostat from the sampler if needed, see Substituir o termostato da amostra.
base-id: 3714788747
id: 27021601479011723