To protect optimal functionality of your flow cell:
G5615-60022 (Cela de fluxo padrão bio-inert, 10 mm, 13 µL, 120 bar (12 MPa) para MWD/DAD, inclui capilar de PEEK 0890-1763 – 0,18 x 1500 mm e conexões PEEK – 5063-6591) (PEEK, ceramic, sapphire, PTFE) for 1260 Infinity III Diode Array Detectors (G7115A):
The recommended pH range of the cell is 1 – 13 (short term 14)
G5615-60005 (Cela de fluxo biocompatível, 8 µL, 20 bar) , (PEEK, fused silica, PTFE) for 1260 Infinity III Fluorescence Detector (G7121A/B)
The recommended pH range of the cell is 1 – 12 (solvent dependent).
If the flow cell is transported while temperatures are below 5 °C, it must be ensured that the cell is filled with alcohol to avoid damage by freezing water.
Aqueous solvents in the flow cell can build up algae. Therefore, do not leave aqueous solvents sitting in the flow cell. Add a small percentage of organic solvents (for example, about 5 % of acetonitrile or methanol).