Quando
|
Ferramentas necessárias
Quant. | p/n | Descrição | |
|---|---|---|---|
1 | ![]()
|
| Chave, 14 mm |
1 | ![]() | 5067-5688 | Chave de torque 1 a 25 Nm com chave de 14 mm |
Peças necessárias
Quant. | p/n | Descrição | |
|---|---|---|---|
1 | ![]() | G7131-60022 | Inlet Valve Biocompatible Flexible Pump |
Pré-requisitos
|
With a 14 mm wrench, unscrew the inlet valve and remove it.
Install the new inlet valve and tighten it using a torque wrench with a 14 mm bit set to approximately 10 Nm.
Insert the inlet weaver, see Replace the Inlet Weaver.
Purge and condition the system to remove air.
base-id: 4133764107
id: 18014402643246091