A functional solvent filter is essential for a good pump performance and for protecting the LC system.
Quando
|
Peças necessárias
Quant. | p/n | Descrição | |
|---|---|---|---|
1 | ![]() | 5041-2168 | Filtro de entrada de solvente (vidro), 20 µm |
See Bottle Head Assembly for related parts.
CUIDADO
Small particles can permanently block the capillaries and valves of the module.
Damage of the module.
NOTA
If the filter is in good condition the solvent will freely drip out of the solvent tube (hydrostatic pressure). If the solvent filter is partly blocked only very little solvent will drip out of the solvent tube.
Remove the solvent filter from the inlet filter adapter and replace it by a new one.
base-id: 3821910923
id: 18014402331392907