エラー ID: 2049
The protection circuit of SSV has switched OFF the valve. During initialization, the processor checks the presence of the SSV connector. If a defective valve or a wrong connection is detected the error message is generated.
考えられる原因 | 対策 | |
|---|---|---|
1 | Defective Solvent Selection Valve. | Change the Solvent Selection Valve. (For G7110B Isocratic pump with SSV only) |
2 | Defective mainboard. | Agilent サービス担当者までお問い合わせください。 |