エラー ID: 2243, 8016
The pump board gets no or wrong pressure signals from the built-in degasser.
Parameters:
0: ADC failure
1: ADC cannot convert pressure signal
2: Measured pressure out of range (negative or higher than 1021 hPa)
The pump board gets no or wrong pressure signals from the built-in degasser:
No valid pressure signal is shown during startup of the degasser
Measured pressure is higher than 32000 hPa
考えられる原因 | 対策 | |
|---|---|---|
1 | Degasser board defective, missing, or not connected to the pump mainboard. | Agilent サービス担当者までお問い合わせください。 |
2 | Degasser sensor defect or not connected to degasser board | Agilent サービス担当者までお問い合わせください。 |
3 | If parameter is 0 and the LED on the degasser board blinks, it is probably because of a defective pump mainboard. | Agilent サービス担当者までお問い合わせください。 |
4 | Degasser sensor defect. | Agilent サービス担当者までお問い合わせください。 |
5 | Degasser sensor not connected to mainboard. | Agilent サービス担当者までお問い合わせください。 |
base-id: 3821425803
id: 9007203076166795