エラー ID: 2040
(MCGV/4CGV)
Valve 0 Failed: valve A
Valve 1 Failed: valve B
Valve 2 Failed: valve C
Valve 3 Failed: valve D
One of the valves of the multi-channel gradient valve has failed to switch correctly.
The processor monitors the valve voltage before and after each switching cycle. If the voltages are outside expected limits, the error message is generated.
Valve 0 Failed: valve A1
Valve 1 Failed: valve A2
Valve 2 Failed: valve B2
Valve 3 Failed: valve B1
One of the solvent selection valves in the module failed to switch correctly.
The processor monitors the valve voltage before and after each switching cycle. If the voltages are outside expected limits, the error message is generated.
考えられる原因 | 対策 | |
|---|---|---|
1 | Gradient valve disconnected. | Ensure the gradient valve is connected correctly. |
2 | Solvent selection valve disconnected. | Agilent サービス担当者までお問い合わせください。 |
3 | Connection cable (inside instrument) not connected. | Agilent サービス担当者までお問い合わせください。 |
4 | Connection cable (inside instrument) defective. | Agilent サービス担当者までお問い合わせください。 |
5 | Gradient valve defective. | Exchange the gradient valve. |
6 | Solvent selection valve defective. | Exchange the solvent selection valve. |