エラー ID: 2207, 2217, 25120
Initialization Failed: Pump channel A
B: Initialization Failed: Pump channel B
The module failed to initialize successfully within the maximum time window.
A maximum time is assigned for the complete pump-initialization cycle. If the time is exceeded before initialization is complete, the error message is generated.
考えられる原因 | 対策 | |
|---|---|---|
1 | Blocked (passive or active) inlet valve. | Exchange the (passive or active) inlet valve. |
2 | Defective pump drive assembly. | Agilent サービス担当者までお問い合わせください。 |
3 | Defective mainboard. | Agilent サービス担当者までお問い合わせください。 |
base-id: 3821268491