エラー ID: 2205, 2215, 2505
Index Missing: Pump channel A
B: Index Missing: Pump channel B
The encoder index position in the module was not found during initialization.
During initialization, the first piston is moved to the mechanical stop. After reaching the mechanical stop, the piston reverses direction until the encoder index position is reached. If the index position is not recognized within a defined time, the error message is generated.
考えられる原因 | 対策 | |
|---|---|---|
1 | Disconnected or defective encoder cable. | Agilent サービス担当者までお問い合わせください。 |
2 | Defective pump drive assembly. | Agilent サービス担当者までお問い合わせください。 |
base-id: 3821051787