The following pages describe maintenance (simple repairs) of the detector that can be carried out without opening the main cover.
Procedure | Typical Frequency | Notes |
|---|---|---|
|
Cleaning of module |
If required |
|
|
Deuterium lamp exchange |
If noise and/or drift exceeds your application limits or lamp does not ignite. |
A wavelength calibration test and an intensity test should be performed after replacement. |
|
Flow cell exchange |
If leaking or if intensity drops due to contaminated flow cell. |
A wavelength calibration test should be performed after replacement. |
|
Leak sensor drying |
If leak has occurred. |
Check for leaks. |
|
Leak handling System replacement |
If broken or corroded. |
Check for leaks. |