エラー ID: 1037
The calibration information needed for your detector to operate correctly has been lost.
During calibration of the detector the calibration values are stored in ROM. If no data is available in the spectrometer ROM, the error message is generated.
考えられる原因 | 対策 | |
|---|---|---|
1 | The detector is new. | Recalibrate the detector. |
2 | The detector has been repaired. | Agilent サービス担当者までお問い合わせください。 |