実行するタイミング
|
必要なツール
数量 | 部品番号 | 説明 | |
|---|---|---|---|
1 | ![]()
|
| ドライバ、Pozidriv #1 PT3 |
1 | ![]()
|
| Screwdriver, flat blade |
2 | ![]()
|
| レンチ、1/4 インチ for capillary connections |
2 | ![]()
|
| ピンセット |
必要な部品
数量 | 部品番号 | 説明 | |
|---|---|---|---|
1 | ![]() | 79880-22711 | Holmium oxide filter |
前提条件
|
注記
See also HOX2 フィルターの適合宣言書.
The glass tends to build a film on its surface even under normal environmental conditions. This is a phenomenon, which can be found also on the surface of several other glasses and has something to do with the composition of the glass. There is no indication, that the film has an influence on the measurement. Even in the case of a thick film, which scatters the light remarkably, no shift of the peak positions is to be expected. A slight change in the absorbance might be possible. Other components within the light path (lenses, windows, ...) are also changing their behavior over the time.
注記
Do not scratch the holmium oxide filter.
The holmium oxide filter can be cleaned with alcohol and a lint-free cloth.
注記
The holmium oxide filter can be cleaned with alcohol and a lint-free cloth.
Perform a holmium oxide test (see Holmium Oxide Test) to check the proper function of the holmium oxide filter.
Install the flow cell, see フローセルの取り外し/取り付け.
Turn on the flow.