エラー ID: 83
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak sensor current to change within defined limits. If the current falls outside the lower limit, the error message is generated.
考えられる原因 | 対策 | |
|---|---|---|
1 | Leak sensor not connected to the mainboard. | Agilent サービス担当者までお問い合わせください。 |
2 | Leak sensor not connected to the on/off switch board. | Agilent サービス担当者までお問い合わせください。 |
3 | Defective leak sensor. | Agilent サービス担当者までお問い合わせください。 |
4 | Leak sensor incorrectly routed, being pinched by a metal component. | Agilent サービス担当者までお問い合わせください。 |
5 | On/Off switch assembly defective. | Agilent サービス担当者までお問い合わせください。 |