エラー ID: 22031
The target pressure of the binary pump degasser has not been reached within the expected time.
Parameter: Pressure in mbar
考えられる原因 | 対策 | |
|---|---|---|
1 | Condensation in degasser chamber due to temperature fluctuation. | Equilibrate and restart module. Use Evacuation Mode if necessary. |
2 | Degasser is defect. | Agilent サービス担当者までお問い合わせください。 |