エラー ID: 22434
The current of the secondary pump drive is too high.
Parameter: 2 or 3 referring to pump drive
考えられる原因 | 対策 | |
|---|---|---|
1 | Blockage of flow path between secondary pump head and pressure sensor, e.g. of the heat exchanger. | Check for blockages in the flow path. Agilent サービス担当者までお問い合わせください。 |
2 | Secondary pump drive is defect. | Agilent サービス担当者までお問い合わせください。 |