エラー ID: 22420
Reading the purge valve tag failed.
Parameter: None
考えられる原因 | 対策 | |
|---|---|---|
1 | Reading of purge valve tag failed. | Check cable connection. |
2 | Purge valve head tag defect or empty. | Replace purge valve head. |
3 | Purge valve tag reader is defect. | Agilent サービス担当者までお問い合わせください。 |
base-id: 4202590859