エラー ID: 22411, 29202
The current consumption of the pump drive is too high.
Parameter: 1 – 4 referring to pump drive
考えられる原因 | 対策 | |
|---|---|---|
1 | Blockage of system before pressure sensor. | Check for blockage of e.g. outlet valve filter frit, purge valve, heat exchanger. |
2 | Drive motor defect. | Agilent サービス担当者までお問い合わせください。 |
base-id: 4202538635
id: 4202538635