エラー ID: 29240
Reading the multipurpose valve tag failed.
Parameter: None
考えられる原因 | 対策 | |
|---|---|---|
1 | Reading of multipurpose valve tag failed. | Check cable connection. |
2 | Multipurpose valve head tag defect or empty. | Replace multipurpose valve head. |
3 | Multipurpose valve tag reader is defect. | Agilent サービス担当者までお問い合わせください。 |