The solvent filters are located on the low-pressure side of the binary pump. A blocked filter therefore does not necessarily affect the high pressure readings of the pump. The pressure readings cannot be used to check whether the filters are blocked or not. If the solvent cabinet is placed on top of the binary pump, the filter condition can be checked in the following way:
Remove the solvent inlet tube from the inlet port of the solvent selection valve or the degasser. If the filter is in good condition, the solvent will freely drip out of the solvent tube (due to hydrostatic pressure). If the solvent filter is partly blocked only very little solvent will drip out of the solvent tube.
警告
キャピラリーまたはチューブの継ぎ手を開けると、溶媒が漏れ出す可能性があります。
毒性や危険性のある溶媒と試薬の取り扱いには、健康上のリスクがある恐れがあります。