エラー ID: 2049
Valve Fuse 0: Channels A1 and A2
Valve Fuse 1: Channels B1 and B2
One of the solvent-selection valves in the module has drawn excessive current causing the selection-valve electronic fuse to open.
考えられる原因 | 対策 | |
|---|---|---|
1 | Defective solvent selection valve. | Restart the pump. If the error message appears again, exchange the solvent selection valve. |
2 | Defective connection cable (front panel to main board). | Agilent サービス担当者までお問い合わせください。 |
3 | Defective mainboard. | Agilent サービス担当者までお問い合わせください。 |
base-id: 3836597003