注記
If the test fails with Max-Light Cartridge Test Cell and new UV-lamp, the optical unit must be replaced.
考えられる原因 | 対策 | |
|---|---|---|
1 | Absorbing solvent or air bubble in Max-Light Cartridge Cell. | Repeat calibration with Max-Light Cartridge Test Cell and compare results. |
2 | Dirty or contaminated Max-Light Cartridge Cell. | Ensure the Max-Light Cartridge Cell is filled with water. Recalibrate. |
3 | UV ランプが古い。 | UV ランプを交換します。 |
4 | Dirty or contaminated optical components. | Run the Cell Test. If the test fails, flush the flow cell. See also Clean the Max-Light Cartridge Cell. |