Holmium Oxide Test evaluation
考えられる原因 | 対策 | |
|---|---|---|
1 | Absorbing solvent or air bubble in flow cell. | 確実にフローセルが水で満たされ、気泡がないようにします。 |
2 | Incorrect calibration | Recalibrate (see Wavelength Calibration) and repeat the test. |
3 | Dirty or contaminated flow cell. | Run the cell test (see Cell Test). If the test fails, exchange the flow cell windows. |
4 | Dirty or contaminated optical components (achromat, windows). | Clean optical components with alcohol and lint-free cloth or replace the parts (see Intensity Test). |
5 | Old or non-Agilent lamp. | UV ランプを交換します。 |