The self-test runs a series of individual tests (described on the next pages), and evaluates the results automatically. The following tests are run:
Filter Test
Slit Test
Dark Current Test
Intensity Test
Wavelength Verification Test
Holmium Oxide Test
Spectral Flatness Test
ASTM Noise Test (without testing the Drift)
実行するタイミング
|
必要な部品
数量 | 部品番号 | 説明 | |
|---|---|---|---|
1 | ![]()
|
| Removed Flow Cell |
前提条件
|
Run the Self Test with Agilent Lab Advisor (for further information see Online-Help of user interface).

Under the tab Signals you can find the detailed signals from the tests.
base-id: 3358400779
id: 3358400779