エラー ID: 30717
The system is in an error state because the calibration of the analog temperature sensor has failed.
考えられる原因 | 対策 | |
|---|---|---|
1 | Sampler incompatibility. | If the hosting sampler is a Vialsampler, verify its compatibility with the Sample Cooler installed. Units with the serial number DEBAT02000 or below are equipped with an analog temperature sensor that is not compatible with the Vialsampler. |
2 | Potential hardware error. | Agilent サービス担当者までお問い合わせください。 |